The ATR—BR is a compact laboratory refractometer for measuring liquids
from low to high viscosity, independent of opacity and colour. The ATR instrument is available in a variety of configurations for laboratory and production quality control.
Main specifications
Models | ATR-BR |
Measuring scales | Refractive Index (RI) Sucrose (%Brix) Standard scales available, Customised scales on request |
Measuring range | 1,32000 – 1,54000 RI / 0 – 100 %Brix |
Resolution | 0,00001 RI / 0,01 %Brix |
Precision | ± 0,00002 RI / ± 0,01 %Brix (measurement range 1,32 – 1,44) and ± 0,00003 RI / ± 0,01 %Brix (measurement range 1,44 – 1,54) |
Reproducibility | 0,00001 RI / 0,01 %Brix |
Measurement mode | Single sample |
Temperature | |
Ambient temperature | +10 – +40 °C |
Temperature measurement | NTC sensor for measurement of sample temperature placed inside the prism |
Temperature control sample | No temperature regulation |
Temperature control range | No temperature regulation |
Temperature resolution | 0.01 °C |
Temperature precision | ± 0.03 °C |
Temperature reproducibility | ± 0.02 °C |
Temperature compensation Brix | +5 – +50 °C |
Additional Information | |
Prism | Saphire |
Light source/wavelength | LED, interference filter 589 nm |
Display | back -lit LCD, 16 x16 characters |
Operation | 20 key membrane including function keys |
Interfaces/communication | RS232 C serial bi-directional port, parallel interface (Centronics), USB optional |
Conformity | International Pharmacopoeia, ASTM, ICUMSA and others |
* Specific standards depend on application and country. Details on request.